Publication Type : Conference Paper
Thematic Areas : Wireless Network and Application
Publisher : 13th International Conference on Electromagnetic Interference and Compatibility (INCEMIC), Andhra University.
Source : 13th International Conference on Electromagnetic Interference and Compatibility (INCEMIC), Andhra University, 2015.
Campus : Amritapuri
School : School of Engineering
Center : Amrita Center for Wireless Networks and Applications (AmritaWNA)
Department : Electronics and Communication
Year : 2015
Abstract : The International Electrotechnical Commission(IEC) IEC 61000-4-2 standard is commonly followed for carrying out the Electrostatic Discharge (ESD) test. An ESD coaxial current target-attenuator-cable chain is specified in the IEC 61000-4-2 edition 2 standard, to measure the discharge currents of ESD generators. In this paper, a method for the calibration of low-frequency transfer impedance of ESD target-attenuator-cable measurement chain is proposed. The proposed method is based on insertion loss measurement of the measurement chain using a vector network analyzer (VNA). Measurement results obtained by proposed insertion loss method and low frequency transfer impedance method as per IEC 61000-4-2 Ed. 2 standard are compared and presented in this paper.
Cite this Research Publication : S. K., “Calibration of Transfer Impedance on ESD Coaxial Target Chain”, in 13th International Conference on Electromagnetic Interference and Compatibility (INCEMIC), Andhra University, 2015.